Guidance on efficacy requirements for biocides: Planning your test product and testing

On this page

Formulation-specific testing

The efficacy information you include in your application should:

The method you use to test efficacy should be relevant to the:

For biocides that come in different dilutions, you may use the same in-use formulation and methods of application to develop efficacy information to support your application.

Replicating the formulation

Your test formulation should be representative of your biocide. For example, it should have the same:

Ingredients

The ingredients and shelf life used in the test sample should be the same as to those in the proposed biocide. The only differences that we accept without confirmatory data are differences in dyes and fragrances under or equal to 1% w/w. Other alternative formulations should be supported by confirmatory efficacy data.

The test product you use to test efficacy for each sample or batch should be formulated at the lower certified limits (LCL) for all representative microorganisms within each microbial category. LCLs (described in Table 1) are calculated using the biocide's labelled nominal active ingredient concentrations.

As there are potential difficulties associated with generating test samples exactly at the LCL, we accept testing up to, and not above, its upper limit.

Table 1: Lower certified limits (LCL)
Nominal concentration (N) of active ingredients LCL Upper limit for LCL
N less than or equal to 1% N minus 10% N LCL plus 2% LCL
1% less than N less than or equal to 20% N minus 5% N LCL plus 1% LCL
20% less than N less than or equal to 100% N minus 3% N LCL plus 0.6% LCL

Testing against representative test organisms in each microbial category should be done using your biocide's LCL. If you want your biocide to have a narrower LCL (approved by us), you should provide quality and stability information that supports the proposed limits.

Refer to quality requirements for biocides for more information.

Additional microorganisms that are not considered representative microorganisms for a microbial class can be tested at or below the nominal concentration. Refer to Table 2 for examples of:

For more information on efficacy data requirements for other microbial categories, refer to the following sections in this guidance:

Table 2: Examples of representative microorganisms that require testing at LCL
Microbial categories on hard or non-porous surfaces Representative test organisms to be tested at LCL

Bacteria disinfection

S. aureus (ATCC 6538) and 1 of either P. aeruginosa (ATCC 15442)Footnote 1 or S. enterica (ATCC 10708)

All other non-spore forming bacteria disinfection claims can be tested at or below nominal concentration

P. aeruginosa should be tested at LCL for hospital disinfectantsFootnote 1

Broad-spectrum virucideFootnote 2

Testing against any small non-enveloped virus from the following families:

  • Picornaviridae (such as rhinovirus)
  • Parvoviridae (such as canine parvovirus)
  • Caliciviridae (such as feline calicivirus)
  • Astroviridae (such as human astrovirus)
  • Polyomaviridae (such as human polyomavirus)

Viruses

Hardest-to-kill virus on the labelFootnote 3

All other virus disinfection claims can be tested at or below the nominal concentration

Fungicide

T. interdigitale (ATCC 9533)

Specific fungi

If your biocide is a general fungicide, then testing at LCL is only required for T. interdigitale

Except for C. auris, additional fungal claims can be tested at nominal concentration

In the absence of a general fungicide claim, specific fungi claims should be tested at the LCL

Food contact surface sanitization

S. aureus (ATCC 6538) and E. coli (ATCC 11229)for non-halide biocides

S. aureus (ATCC 6538) or S. enterica (ATCC 6539)for halide biocides

All other food contact surface sanitization claims can be tested at or below nominal concentration

Non-food contact surface sanitization

S. aureus (ATCC 6538) and 1 of either

K. pneumoniae (ATCC 4352) or K. aerogenes (ATCC 13048)

All other non-food contact surface sanitization claims can be tested at or below nominal concentration

Mycobactericide

M. bovis (ATCC 35743) or M. terrae (ATCC 15755)

Specific mycobacteria

All specific mycobacteria

If there's no general mycobactericide claim, all specific mycobacteria claims should be tested at the LCL

Sporicide

B. subtilis (ATCC 19659) and C. sporogenes (ATCC 3584)

Specific spores

All spore claims should be tested at the LCL

Footnote 1

ATCC: American Type Culture Collection
P. aeruginosa is required for hospital disinfectants

Return to footnote 1 referrer

Footnote 2

Previous Health Canada guidance allowed the following viruses to be used as qualifying surrogate viruses for a broad-spectrum virucide claim:

  • poliovirus type 1 (ATCC VR-1562 Chat Strain)
  • human adenovirus type 5 (ATCC VR-5)
  • bovine parvovirus (ATCC VR-767)
  • canine parvovirus (ATCC VR-2017)

All broad-spectrum virucide claims approved for a surface disinfectant under the Food and Drug Regulations using 1 of these viruses will be acceptable under the Biocides Regulations. Learn more about our approach for emerging viral pathogens. We no longer recommend the use of poliovirus as a qualifying surrogate virus in support of a broad-spectrum virucide or emerging viral pathogen claim for surface disinfectants. This aligns with Canada's commitment to the World Health Organization (WHO)'s global polio eradication initiative.

Return to footnote 2 referrer

Footnote 3

The hardest-to-kill virus is any small, non-enveloped virus noted on the label. Otherwise, it's a large-non-enveloped virus. If there are no large-non-enveloped viruses, it's any of the tested enveloped viruses.

Return to footnote 3 referrer

You should provide a certificate of analysis for the tested batches to demonstrate that testing has been done at the appropriate concentration. If you use a method that cannot differentiate between active ingredients in a formulation, you should submit analytical testing for the combined ingredients (for example, quaternary ammonium compounds (QACs)). The LCL can be determined for the sum of these ingredients.

To test your biocide at the LCL, you may either produce a test batch with an active ingredient at the LCL or dilute your biocide with water. Biocides that may be reactive or less stable in water may be diluted with a primary solvent already present in the formulation. Avoid using emulsifiers or surfactants as diluents even if they're present in the formulation, as these may alter the biocide's efficacy.

Biocides in a solid physical form (powder or tablets) should be diluted with more water than indicated on the label to achieve an in-use concentration at LCL. For example, before efficacy testing:

Sometimes it may not be feasible to test a biocide formulated at the exact LCL. If this is the case, provide a rationale in your efficacy summary to support why your efficacy testing was not conducted at the exact lower limits.

If it's not possible to dilute the biocide, you may choose other ways to test at the LCL, for example, testing:

For more information on acceptable accelerated stability conditions, consult:

Diluent or activator use

If your biocide is to be diluted or activated (or both) before using, your test formulation should be representative of the biocide's labelled directions for use. This means it should, for example:

All microorganisms in the same microbial category (bacteria, viruses, fungi) should be tested at the same level of water hardness. Hard water tolerance levels may differ with the level of antimicrobial activity claimed for a biocide. Unless your biocide label says otherwise, we recommend using water with a minimum hardness of 200 ppm (parts per million), expressed as the amount of calcium carbonate (CaCO3) present in your testing.

If your biocide indicates efficacy claims specifically in hard water for any of the microbial categories, testing should be conducted with water at a higher hardness level for all microorganisms within that microbial category. For example, if your biocide makes claims of food contact surface sanitization efficacy in 500 ppm (parts per million) of hard water, all microorganisms for that claim should be tested with 500 ppm water.

We may allow you to use distilled water as a diluent in your test product if you provide an acceptable justification. An example would be if the devices used in testing are sensitive to calcium carbonate buildup. If this is the case, your biocide label should indicate to dilute with distilled water.

Replicating the physical form and method of application

The method of application and conditions of use for your test product and your biocide should be the same. This applies to biocides with the following physical forms and methods of application:

Liquid biocides

Immersion, mopping or using a cloth or sponge are methods of application that involve evenly applying a sufficient amount of a liquid biocide to a surface. Biocides with these methods of application can be tested using immersion methods, such as the "use dilution method" for bacteria.

Liquid biocides that are intended to be sprayed on a surface should be using spray methods such as the "germicidal spray test" for bacteria.

Non-volatile biocides are expected to evenly cover a treated surface for the entire contact time. Biocides that have both spray and non-spray applications may be tested using an immersion method.

Volatile biocides (containing isopropyl alcohol or ethanol as an active ingredient, for example) should be tested using spray methods. This is because immersion methods do not accurately simulate the way in which volatile biocides perform on environmental surfaces.

Conduct a spray test on your biocide. If it evenly covers the surface of the carrier for the entire contact time, then a spray method can be used for efficacy testing.

Towelette biocides

If your biocide is a single-use pre-saturated or impregnated towelette, you should use the towelette to test the biocide's efficacy. Your test method should include detailed procedures for:

You can also use the liquid expressed from a towelette to conduct your testing. In your application include analytical data on the concentration levels of the active ingredients in the expressed liquid.

To test with a towelette, use 1 towelette to treat either:

Each carrier should be wiped back and forth up to 3 times, for a total of 6 passes, before moving to the next carrier. A pass is defined as moving from 1 side of the carrier to the other with a single motion. The contact time begins after 6 passes.
For disinfectant claims against bacteria, mycobacteria and fungi, we recommend either of the following test methods:

We recommend the ASTM E1053 method for virucidal claims. Use 1 towelette to wipe 1 test carrier.

You can bridge data from a liquid biocide for the same biocide in towelette form using confirmatory data. All representative microorganisms in each microbial category should be tested using the biocide in towelette form. If the towelette biocide has the same efficacy claims as the biocide in liquid form, you can bridge data if you:

Extended-use and reusable biocides

You should provide data to support your biocide's effectiveness at the end of its shelf life:

For example, a biocide that is used for a certain number of days for immersion application after repeated soaking of contaminated instruments.
The data should reflect simulated use, soiling and dilution as applicable.

Vapour or gas biocides

If your biocide is a vapour, mist or gas, you should test your biocide under similar conditions and settings (for example, simulated-use testing). This includes the type and size of the areas the biocide is for use in. Examples of vapour or gas biocides include those:

When testing vapour or gas biocides, be sure that all key parameters for efficacy are monitored throughout the use area and reflect the labelled conditions, such as:

When planning your testing, you should:

For guidance on the test method for vapour or gas biocides, consult:

Testing efficacy

In addition to replicating your biocide's formulation, physical form and conditions of use, you should use the same conditions and parameters as the referenced test method (unless test method does not prescribe the conditions and parameters). Conditions and parameters include:

The number of carriers should be based on the most recent version of the test method being referenced.

All microorganisms should be tested in similar conditions to how the biocide will be used. If the biocide will not be used in a conventional way, test modifications or alternative test methods may be needed to support the efficacy of the biocide for its proposed use. Contact us to discuss the appropriate alternative test methods or modifications for your biocide.

Contact time

Your efficacy testing should use the same (or a shorter) contact time as the time indicated on your proposed biocide label.

Note: Any change in contact time in efficacy testing may be restricted by the limitations of the test method or the proposed method of application. For instance, an exposure period greater than 10 minutes for a biocide that will likely evaporate from a treated hard surface within 10 minutes. Shorter contact times may be required for some uses where the 10-minute contact time is not achievable.

Your biocide should meet the performance standard set for the test method and target microorganism within a 10-minute contact time. Exceptions may be given to disinfectants that direct users to immerse objects in the solution for a specified time.
You will not be able to propose contact times above 10 minutes unless either:

Contact us for a pre-submission meeting if you require a longer contact time.
For liquid or spray biocides with volatile active ingredients, you can determine the maximum contact time by testing your biocide's evaporation over the proposed contact period using a wetness test.

For spray-only biocides, do not use a test method that immerses your test organism in the biocide fluid. Immersion does not accurately represent the way biocides perform on surfaces. Instead, use 1 of the following test methods:

You may use an immersion test method for non-volatile biocides that are applied by both spray and non-spray methods. Volatile formulations should be tested using spray methods as this accurately represents the way volatile biocides perform on surfaces.

For biocides in towelette forms, you may use a wetness test to determine the maximum contact time. This test will:

Temperature

Your efficacy testing should have the same temperature conditions as those indicated your biocide's label. Otherwise, conduct your test between 18°C and 25°C, including the steps to:

Neutralization

Neutralization inactivates your test product's antimicrobial activity. For all efficacy testing, follow procedures to neutralize your biocide at the end of all contact times. This will prevent residual effects in the subculture medium.

Confirm how you will neutralize the product when you test for efficacy or as required by the test method. The neutralization procedure should align with your test method.
You may neutralize your biocide by:

If you are unable to validate the neutralizers used for all biocide tests against microorganisms (except for viruses) using standard test methods, refer to the:

Organic burden

If your biocide will be used in light to moderate amounts of soil, your organic soil load should be:

Examples of these types of biocides are:

Biocides for use on soft surfaces and for treating an entire room require a 5% soil load. For these applications, we recommend you use:

Acceptable test methods

In general, Health Canada recognizes biocide test methods and protocols published by international standards organizations or other regulators. You should use the current official version of the test method.

You may not apply requirements from different test methods unless you have a suitable rationale for doing so. An example would be modifying test methods commonly recommended by other regulators or international organizations to test alternate organisms or product forms.

You should follow all prescribed information outlined for the test method you use. If the method does not outline certain information (for example, microbial count or performance criteria), you should meet the requirements outlined in this guidance.
In general, we recognize test methods published by:

Note: When using European (EN) test methods, phase 1 test methods are only considered acceptable as secondary supporting evidence of efficacy. While they cannot be used solely to support a proposed claim and are not accepted for biocide authorization, they can be used to help develop a proposed biocide. You should conduct all testing using at least 2 batches and phase 2 test methods.

Sourcing test organisms

Your efficacy testing for each microbial category claimed on your biocide's label should use representative test organisms.

We recommend that you use test organisms available for purchase through the American Type Culture Collection (ATCC). You may use microorganism stocks purchased from other suppliers if they:

Batch replication, microbial counts and performance standards

To show that efficacy results are reproducible and statistically sound, you should have:

Find more information, consult:

The microbial count of a sample (inoculum counts or carrier counts) determines the microbial challenge level for the test. For a test result to be valid, the microbial levels should be within the prescribed limits for the test method.

The way you determine the microbial count should be in line with the test method. For suspension test methods, you should determine the microbial count of the sample. For carrier test methods, you should determine the microbial counts on untreated dried control carriers after applying the sample.

In general, it's acceptable if the microbial counts exceed the prescribed levels outlined in the test requirements as long as the biocide meets the prescribed performance criteria.

Re-testing

A biocide may fail to meet the performance standard required for the efficacy claim due to contamination or not meeting the prescribed carrier count.

If a laboratory test indicates that the failure is due to a false positive, you may re-test that particular batch.

If you decide to re-test, you do not need to use:

You may conduct up to 2 re-tests for a contamination or population control failure. You do not need to re-test for carrier population controls that are greater than the acceptance criteria (unless otherwise specified in the test methods).

If the issue has not been resolved after 2 re-tests, we will not accept the efficacy information.

When re-testing, conditions must be identical to the original test.

Failure due to contamination

Contaminants are microorganisms that are not the test organism in the study. Contamination in the test system may invalidate the assay. We recommend that you follow the US EPA's 810.2000 product performance test guideline to help you determine whether you need to retest the biocide and if so, how.

Use 1 of the following methods to identify the contaminant and presence or absence of the test organism:

If contamination happens again, the laboratory may consider a quality plan to correct the issue. Quality plans can include mitigation strategies such as:

Failed re-testing

If you re-tested and your results continue to fail to meet performance standards, you may wish to do a different test by changing the test conditions. This will affect your labelled claim for your biocide. Examples of changes include:

Confirmatory data requirements

You may use confirmatory data with reduced batch replication requirements to support situations that may affect your biocide's efficacy, such as:

We accept reduced batch replication requirements for confirmatory data for:

For more information, visit:

Formulation variations

If you include formulation variations in your application, make sure they are substantially similar to the basic formulation.

All variations should be supported by a single benefits (efficacy) package and label text to demonstrate that they:

When proposing formulation variations, you should include both:

The confirmatory data that you include in the application should:

For more information, visit the:

You do not need to provide confirmatory efficacy data when:

For more information, consult:

Towelette biocides

Confirmatory testing allows you to bridge additional specific claims approved for your liquid biocide to the same biocide in towelette form. This confirmatory data should include the representative test organisms for each microbial category claimed on the label of your towelette biocide and be tested using towelettes. If acceptable, you do not need efficacy data to support all other additional claims.

When conducting confirmatory testing, you should:

Page details

Date modified: